1-10
General information
TRI-BAND ELECTRIC FIELD PROBE EP-3B-01
Wide band
Low pass
High pass
Frequency range
0.1 – 3000 MHz
0.1 – 862 MHz
933 – 3000 MHz
Reading range
0.2 – 200 V/m
Overload
600 V/m
Dynamic range
> 60 dB
Linearity
± 0.5 dB (0.5 to 100 V/m)
@ 50 MHz
± 0.5 dB (0.5 to 100 V/m)
± 0.5 dB (0.5 to 100 V/m)
Resolution
0.01 V/m
Sensitivity
0.2 V/m
Flatness @ 20V/m
1 – 200 MHz ±0.8 dB
150 kHz - 3 GHz ±1.5 dB
1 – 200 MHz ±0.8 dB
150 kHz - 862 MHz ±1.5 dB
933 – 3000 MHz ±1.5 dB
Anisotropy @ 6V/m
± 0.8 dB @ 50 MHz
(typical 0.6 dB)
+/- 0.8 dB @ 1 GHz
(typical 0.6 dB)
Attenuation out of band
Not applicable
933 MHz – 3 GHz > 23 dB
(respect to 50 MHz)
0.1 – 862 MHz > 23 dB
(respect to 1 GHz)
Rejection of magnetic
field
> 20 dB
Temperature error
0.1 dB/°C
A/D conversion
On board
Calibration factors
On board E
2
prom
Temperature sensor
On board
Dimensions
Length 450mm, diameter 55mm
Weight
180g
Fig. 1-5
EP-3B-01 Probe