![multicomp pro MP720021 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/multicomp-pro/mp720021/mp720021_manual_1854297053.webp)
User Manual
48
Configure
Idle Time
Set idle time before the edge counting in Nth Edge
Trigger. Adjust M rotary control or click
to set
idle time press
panel button or click
to move cursor to choose which digit to be set.
Available range is 30ns-10s, default 100ns.
Edge Num
Set the edge number value of
“
N
”
in Nth Edge trigger.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Reset
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust
M
rotary control or click
to
set time interval before another trigger occur, press
panel button or click
move cursor
to choose which digit to be set.
Set Holdoff time as 100 ns
Logic Trigger
Trigger according to logic relation.
In Logic Trigger mode, the trigger setting information is displayed on bottom right of
the screen, for example,
,
indicates that
trigger type is Logic, logic mode is AND, CH1 high level and trigger level is 0.00mV.
Logic Trigger menu
list:
MENU
SETTING
INSTRUCTION
Mode
Logic
Set vertical channel trigger type as Logic trigger.
Logic
Mode
AND
OR
XNOR
XOR
Set logic mode as AND.
Set logic mode as OR.
Set logic mode as XNOR.
Set logic mode as XOR.
Содержание MP720021
Страница 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Страница 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Страница 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Страница 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Страница 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Страница 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...
Страница 144: ...User Manual 139 7 Persist on P63 It s normal as the Oscilloscope is working hard on many more data points...