User Manual
41
3. Pulse Width Trigger
Pulse trigger occurs according to the width of pulse. The abnormal signals can be
detected through setting up the pulse width condition.
In Pulse Width Trigger mode, the trigger setting information is displayed on bottom
right of the screen, for example,
,
indicates that trigger type
is pulse width, trigger source is CH1, coupling is DC, polarity is positive, and trigger
level is 0.00mV.
Pulse Width Trigger menu
list:
MENU
SETTING
INSTRUCTION
Single Mode
Pulse
Set vertical channel trigger type as pulse trigger.
Source
CH1
CH2
CH3
CH4
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source
Coupling
AC
DC
Not allow DC portion to pass.
Allow all portion pass.
when
Polarity
Choose the polarity
Select pulse width condition and adjust the
M
rotary
control or click
to set time, press
panel button or click
to move cursor to
choose which digit to be set.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Reset
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust
M
rotary control or click
to
set time interval before another trigger occur, press
panel button or click
to move
cursor to choose which digit to be set.
Set Holdoff time as 100 ns
4. Slope Trigger
Slope trigger sets the oscilloscope as the positive/negative slope trigger within the
Содержание MP720021
Страница 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Страница 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Страница 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Страница 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Страница 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Страница 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...
Страница 144: ...User Manual 139 7 Persist on P63 It s normal as the Oscilloscope is working hard on many more data points...