![multicomp pro MP720021 Скачать руководство пользователя страница 147](http://html1.mh-extra.com/html/multicomp-pro/mp720021/mp720021_manual_1854297147.webp)
User Manual
142
Analogue Bandwidth
MP720025
MP720021
100 MHz
Single Bandwidth
MP720025
MP720021
DC to 100 MHz
Low Frequency
≥
10 Hz (at input, AC coupling, -3 dB)
Rise time (at input,
Typical)
MP720025
MP720021
≤ 3.5 ns
DC gain accuracy
MP720025
1 mV
±4%
≥2 mV
±3%
MP720021
1 mV
±3%
≥2 mV
±2%
DC accuracy (average)
Delta Volts between any two averages of
≥16 waveforms acquired with the same
scope setup and ambient conditions (
△
V):
±(3% rdg + 0.05 div)
Waveform inverted ON/OFF
Measurement
Cursor
△
V,
△
T,
△
T&
△
V between cursors,
auto cursor
Automatic
Period, Frequency, Mean, PK-PK, RMS,
Max,
Min,
Top,
Base,
Amplitude,
Overshoot, Preshoot, Rise Time, Fall Time,
+Pulse Width, -Pulse Width, +Duty Cycle,
-Duty Cycle, Delay A→B
, Delay
A→B , Cycle RMS, Cursor RMS, Screen
Duty, Phase A→B
, Phase A→B
, +Pulse
Count, -Pulse Count, Rise Edge Count, Fall
Edge Count, Area, and Cycle Area.
Waveform Math
+
,
-
, *, / ,FFT, FFTrms, Intg, Diff, Sqrt,
User Defined Function, digital filter (low
pass, high pass, band pass, band reject)
Decoding Type
(optional)
RS232, I
2
C, SPI, CAN
Waveform storage
100 waveforms
Lissajous
figure
Bandwidth
Full bandwidth
Phase
difference
±3 degrees
Communication
port
Standard USB Host, USB Device; Trig Out(Pass/Fail); LAN port
Optional VGA port
Printer
Compatibility
PictBridge
Содержание MP720021
Страница 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Страница 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Страница 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Страница 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Страница 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Страница 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...
Страница 144: ...User Manual 139 7 Persist on P63 It s normal as the Oscilloscope is working hard on many more data points...