User Manual
12
compensated or presents a compensation deviation will result in the measuring error
or mistake. For adjusting the probe compensation, please carry out the following
steps:
1.
Set the attenuation coefficient of the probe on the menu as 10X and that of the
switch in the probe as 10X
(see "
How to Set the Probe Attenuation Coefficient
on P12), and connect the probe with the CH1 channel. If a probe hook tip is used,
ensure that it keeps in close touch with the probe. Connect the probe tip with the
signal connector of the probe compensator and connect the reference wire clamp
with the ground wire connector of the probe connector, and then push the
Autoset
button on the front panel.
2.
Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see
and
Overcompensated
Compensated correctly
Under compensated
Figure 3-6 Displayed Waveforms of the Probe Compensation
3.
Repeat the steps mentioned if needed.
Figure 3-7 Adjust Probe
How to Set the Probe Attenuation Coefficient
The probe has several attenuation coefficients, which will influence the vertical scale
factor of the oscilloscope.
To change or check the probe attenuation coefficient on the menu of oscilloscope:
(1)
Push the function menu button of the used channels (
CH1 - CH4
button).
(2)
Select
Probe
in the bottom menu; select
Attenu
in the right menu, turn the
M
rotary control to select the proper value corresponding to the probe.
This setting will be valid all the time before it is changed again.
Содержание MP720021
Страница 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Страница 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Страница 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Страница 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Страница 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Страница 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...
Страница 144: ...User Manual 139 7 Persist on P63 It s normal as the Oscilloscope is working hard on many more data points...