CellScale
®
System • 3750CS-5 User Guide Page 81
MEASUREMENT SYSTEMS INTERNATIONAL
Software V2-21 or greater, 2nd Generation Modem
TO CALIBRATE
Use this Calibration procedure for calibration of the CellScale using the 3750CS. If capacity or resolution changes
are necessary (such as an initial setup of a system) first reset the Calibration with the procedure on page 111. The
CellScale supports up to 10 points of span calibration, however, calibrating with the 3750CS limits the calibration
to 5 span points. Zero and at least 1 span point are required for a successful calibration. The highest value test
weight should be at least 20% of capacity, but 50% or more is preferable.
Before starting the following procedure, you must be in the Top Menu Level of Calibration. See “Enable Calibration”.
1) Use the UP or DOWN scroll key to scroll to the “STD
CAL” message.
2) Press ENTER.
If this is a recalibration, jump to step 9. If this is the initial calibration or
following a Reset Cal procedure, go to step 3.
3) Next set the default units the scale is calibrated in.
Use the UP or DOWN key to scroll to the desired
Calibration Unit.
In this example, we’ll pick “KILOGRAM”. The capacity initially programmed
in the calibration units determines the overload value.
4) Press ENTER to set the Calibration Units.
5) Next set the capacity of the system in Calibration
Units. The display reads “CAPACITY”. Use the
numeric keys to set the capacity value.
In this example we’ll set the meter Capacity at 1000 kg. Capacities range
up to 999,999 calibration units in any units. During capacity entry, you can
back step delete any incorrect entry using the CLR (ZERO) key.
6) Push ENTER to complete the capacity value entry.
7) Next, use the UP or DOWN keys to enter the resolu-
tion in scale divisions (d). The CellScale defaults to
3000 counts or less resolution. The message display
reads “COUNT BY”. The numeric display indicates
the calculated “d” value. Use the UP key to decrease
resolution (increase the count-by). Use the DOWN
key to increase resolution (decrease the count-by).
Increments are in a 1-2-5 sequence.
In this example we’ll change the scale divisions to 0.1 kg
8) Push ENTER when the desired scale division is
displayed.
9) Zero Calibration – The message reads “Ø SCALE”.
Remove all weight from the scale and wait until
the scale is motionless. When the load receptor is
stable push ZERO or ENTER. The display reads
“CAL’ING”. Wait for about 5 seconds. If the detected
zero weight is within acceptable limits the message
“WEIGHT 1” will appear.
1
...
1
STD CAL
2
POUNDS
3
KILOGRAM
4
0
CAPACITY
5
1
5
10
5
100
5
1000
6
COUNTBY
7
7 (1000kg default)
0.5
7
0.2
7
0.1
8
0 SCALE
11
CAL’ING
WEIGHT 1