61
Maintenance
If a sensor is in Diffusion Supervision Fault, the device will still allow the user to attempt to calib-
rate the sensor. In the rare event that a sensor calibration is successful, but the Diffusion Super-
vision system continues to detect a blockage in the gas path, the Diffusion Supervision Fault
will return in approximately 24 hours. In this scenario, there is likely an environmental factor
affecting the Diffusion Supervision detection. Provided the sensor can successfully calibrate,
the sensor is still capable of detecting gas. To prevent the Diffusion Supervision Fault from
reoccurring, a user can either disable Diffusion Supervision in the Sensor Settings or replace
the sensor.
6.4
Troubleshooting
Table 17 below lists all fault messages, their priority levels, and corrective actions required to
resolve them. The faults are listed in alphabetical order. Lower priority messages are output only
after the highest priority message is cleared.
US
Содержание TG5000
Страница 1: ...Operating Manual TG5000 Gas Monitor Order No 10212126 L Y REV 0 US ...
Страница 36: ...36 Operation US ...
Страница 46: ...46 Operation US ...
Страница 62: ...62 Maintenance US ...
Страница 63: ...63 Maintenance US ...
Страница 64: ...64 Maintenance US ...
Страница 65: ...65 Ordering Information Tab 17 Troubleshooting 1 Lower numbers have higher priority US Maintenance ...
Страница 70: ...70 Appendix Calibration Guide and Additional Gases US ...
Страница 71: ...71 Appendix Calibration Guide and Additional Gases R 32 US ...
Страница 72: ...72 Appendix Calibration Guide and Additional Gases US ...
Страница 73: ...73 Appendix Calibration Guide and Additional Gases Tab 21 XIR PLUS Calibration Guide for Additional Gases US ...
Страница 75: ...For local MSA contacts please visit us at MSAsafety com ...