![MSA TG5000 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/msa/tg5000/tg5000_operating-manual_1846413051.webp)
51
Calibration
Tab. 16 Calibration Kits
1
Balance Air
2
Balance Nitrogen
Flow Rates:
CALKIT1 = 1 liter/min
CALKIT2 = 0.25 liter/min
5.2
Calibration Frequency
The frequency of calibration gas testing depends on the operating time, chemical exposure, and
type of sensor. Especially in new installations or applications, it is recommended that the first
sensors be calibrated more often to establish the sensor performance in this particular environ-
ment.
For this, you typically record the "as found" and "as left" gas values and track the percent adjust-
ment over time. Then, gradually extend calibration intervals until the percent adjustment is greater
than the expected accuracy of the sensor.
US
Содержание TG5000
Страница 1: ...Operating Manual TG5000 Gas Monitor Order No 10212126 L Y REV 0 US ...
Страница 36: ...36 Operation US ...
Страница 46: ...46 Operation US ...
Страница 62: ...62 Maintenance US ...
Страница 63: ...63 Maintenance US ...
Страница 64: ...64 Maintenance US ...
Страница 65: ...65 Ordering Information Tab 17 Troubleshooting 1 Lower numbers have higher priority US Maintenance ...
Страница 70: ...70 Appendix Calibration Guide and Additional Gases US ...
Страница 71: ...71 Appendix Calibration Guide and Additional Gases R 32 US ...
Страница 72: ...72 Appendix Calibration Guide and Additional Gases US ...
Страница 73: ...73 Appendix Calibration Guide and Additional Gases Tab 21 XIR PLUS Calibration Guide for Additional Gases US ...
Страница 75: ...For local MSA contacts please visit us at MSAsafety com ...