3.3.9 XP Remote Sensor Duct Mount
Duct mount kits are available for monitoring atmosphere inside flat or round ducts. Round duct
mount kits are available for small ducts 12-20" (305-508 mm) in diameter (Part Number
10179124) and large ducts 20-40" (508-1016 mm) in diameter (Part Number 10179321). The flat
duct mount (Part Number 10176947) is universal for flat ducts.
NOTICE
Consider the sensor type before choosing a duct mount location. ULTIMA XIR Plus sensors
should be mounted horizontally and all other sensors should be mounted vertically.
NOTICE
Air flow in the duct must be zero to ensure proper calibration.
Fig. 16
Flat Duct Mount
Fig. 17
Round Duct Mount