MAINTENANCE MSA
48
ALTAIR 5 - Operating Manual
Figure 4-4
Sensor Replacement
(1) Verify that the instrument is turned OFF.
(2) Remove the battery pack.
(3) Remove the two remaining case screws, and remove the case front.
(4) Gently remove the sensor to be replaced.
•
Using fingers only, gently remove the tox, combustible or oxygen
sensor.
(5) Carefully align the new sensor contact pins with the sockets on the
printed circuit board.
(6) Press the new sensor into place.
•
Ensure groove in combustible sensor is aligned with notch in
sensor slot.
•
If a sensor is not to be installed, ensure that a sensor plug is
installed properly in its place.
(6) If replacing sensor filter at this time:
•
Carefully peel off old filter taking care not to damage the inside of
the case.
•
On new filter peel off backing exposing adhesive. Note proper
orientation of filter and apply to inside of front case adhesive side
against case.
•
Press filter into place taking care not to damage filter surface.
(7) Reinstall the sensor gasket in the case front.
Ensure sensor gasket is properly installed.
(8) Re-install the screws.
(9) Go into the sensor setup and turn ON the sensor. If necessary
calibrate instrument before use.
Allow sensors to stabilize at least 30 minutes before calibration.
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...