MAINTENANCE
ALTAIR 5 - Operating Manual
47
Figure 4-3
Battery Replacement
1 Battery
holder
(3) For alkaline battery packs (ALTAIR 5 only):
•
Remove the battery holder circuit board from the instrument
•
Replace the 3 cells, using only those listed on the label.
Be sure to observe proper polarity on the cells.
•
Place the battery holder circuit board back in the instrument and
reinstall the door.
•
Tighten the 2 screws.
4.4. Live Maintenance Procedure - Replacing or Adding a Sensor
NOTICE
Before handling the PC board, the user must be properly grounded;
otherwise, static charges could damage the electronics. Such damage
is not covered by the warranty. Grounding straps and kits are available
from electronics suppliers.
WARNING
Remove and reinstall sensors carefully, ensuring that the components
are not damaged; otherwise instrument, intrinsic safety may be
adversely affected, wrong readings could occur, and persons relying on
this product for their safety could sustain serious personal injury or
death.
To add a sensor to an instrument that is not already equipped with a full
array of sensors, remove the sensor plug from in front of the formerly
unused sensor housing.
NOTICE
While instrument case is open, do not touch any internal components
with metallic/conductive objects or tools.
Damage to the instrument
can occur.
1
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...