OPERATION
ALTAIR 5 - Operating Manual
29
3.4. Instrument
Setup
The instrument allows the user to access and modify the following
parameters through direct button interface:
•
Calibration Options
•
Alarm Options
•
Instrument Options
These menus can be accessed only from the measure page by pressing
and holding the
and
buttons simultaneously until you are prompted for
a password.
The operation is as follows:
(1) Turn the instrument ON and wait until the measure page appears.
(2) Press the
and
buttons simultaneously.
•
The default password is "672".
Monochrome display
Color display
PASSWORD PASSWORD
000
000
(3)
Enter the first digit by pressing the
or
button and confirm
with the
button.
•
The cursor jumps to the second digit.
(4) Enter the second as well as the third digit.
•
Incorrect password: instrument returns to the Main Page.
•
Correct password: user can set the instrument options.
The password can be changed with a PC through the MSA Link
TM
software.
The following Options are available by pressing the
and
buttons
:
•
Calibration Options - see Section 3.4.1
•
Alarm Options - see Section 3.4.2
•
Instrument Options - see Section 3.4.3
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...