OPERATION
ALTAIR 5 - Operating Manual
37
3.5. Data
Logging
Connecting Instrument to PC
(1) Switch ON the ALTAIR 5 and align the Datalink Communication port
on the ALTAIR 5 to the IR interface of the PC.
(2) Start the MSA
Link
TM
software on the PC and start connection by
clicking the connect icon.
See MSA Link
TM
documentation for detailed instructions.
3.6. Function Tests on the Instrument
Alarm Test
(1) Turn ON the instrument. The user should verify that:
•
alarm LEDs flash
•
horn sounds briefly
•
vibrating alarm triggers briefly.
3.7. Safe LED
The instrument is equipped with a green "SAFE LED". This green
SAFE LED flashes every 15 seconds under the following conditions:
•
the green SAFE LED is enabled
•
instrument is on the normal Measure Gases page
•
combustible reading is 0% LEL or 0%CH
4
•
Oxygen (O
2
) reading is 20.8%
•
Toxic sensor readings are 0 ppm
•
no gas alarms are present (low or high)
•
instrument is not in Low Battery warning or alarm
•
Toxic sensor, STEL and TWA readings are 0 ppm.
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...