US
OPERATION MSA
30
ALTAIR 5 - Operating Manual
3.4.1 Calibration
Options
Monochrome display
Color display
CALIBRATION OPTIONS
CALIBRATION OPTIONS
The Calibration Options menu allows the user to:
•
modify the calibration cylinder settings
(
CYLINDER SETUP
)
•
enable/disable calibration due and to set the number of days
(
CAL DUE OPTIONS
)
•
enable/disable the option to show the last cal date at turn on and
(
LAST CAL DATE
)
•
enable/disable the option for password protected calibration
(
CAL PASSWORD
)
Press
•
the
button go to next page
•
the
button to go previous page
•
the button to enter setup.
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
(1) Press
the
button to enter setup.
•
The screen for the first calibration cylinder displays.
(2) Press
•
the
or
button to change the value.
•
the button to confirm the setup.
With this confirmation the instrument automatically moves to the next
cylinder setting.
(3) Repeat the sequence for changing the required settings for all
necessary gas values.
After the last setting is performed, the instrument returns to the
Calibration Options menu.
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...