ALTAIR 5 - Operating Manual
2.7. Monitoring the Oxygen Concentration
The instrument monitors the oxygen concentration in ambient air. The alarm
setpoints can be set to activate on two different conditions:
•
Enriched - oxygen concentration > 20.8 % or
•
Deficient - oxygen concentration < 19.5 %.
While the instrument can detect up to 25% oxygen in the ambient air, it is
approved for use only up to 21% oxygen-content.
WARNING
If an alarm activates while using the instrument, leave the area
immediately.
Remaining on site under such circumstances can cause serious
personal injury or death.
When the alarm setpoint is reached for either of the above conditions:
•
the alarm message displays and flashes in combination with the
corresponding gas concentration
•
backlight turns on
•
alarm sounds
•
alarm LEDs flash
•
vibrating alarm triggers
The LOW alarm (oxygen deficient) is latching and will not reset when the
O
2
concentration rises above the LOW setpoint. To reset the alarm press
the
button. If the alarm is latching, the
button silences the alarm for five
seconds.
False oxygen alarms can occur due to changes in barometric pressure
(altitude) or extreme changes in ambient temperature.
It is recommended that an oxygen calibration be performed at the
temperature and pressure of use. Be sure that the instrument is in known
fresh air before performing a calibration.
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...