US
MSA
DESCRIPTION
ALTAIR 5 - Operating Manual
23
2.8. Monitoring
Combustible
Gases
The instrument can monitor the concentration of a variety of combustible
gases in ambient air. Which combustible gases are monitored depends on
the installed sensors.
The instrument displays the gas concentration in % LEL or % CH
4
on the
Measuring page until another page is selected or the instrument is turned
OFF.
WARNING
If an alarm activates while using the instrument, leave the area
immediately.
Remaining on site under such circumstances can cause serious personal
injury or death.
The instrument has two alarm setpoints:
•
HIGH Alarm
•
LOW Alarm
If the gas concentration reaches or exceeds the alarm setpoint, the
instrument:
•
backlight turns on
•
displays and flashes the alarm message either LOW alarm or the
HIGH alarm
•
enters an alarm state.
When gas reading exceeds 100% COMB or 5.00% CH
4
, the instrument enters
a Lockalarm
TM
state and displays “XXX” in place of the actual reading. This
state can only be reset by turning the instrument OFF and ON.
WARNING
A combustible gas reading of "XXX” indicates the atmosphere is above
100 % COMB or 5.00 %vol CH
4
and an explosion hazard exists. Move
away from the contaminated area immediately.
In such cases, the instrument LockAlarm feature activates.
Содержание altair 5
Страница 59: ...US MSA ORDER INFORMATION ALTAIR 5 Operating Manual 59 Replacement parts 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 73: ...US MSA APPENDIX FLOW CHARTS ALTAIR 5 Operating Manual 73 ...
Страница 133: ...ES INFORMACIÓN DE PEDIDOS MSA 60 ALTAIR 5 Manual de funcionamiento Piezas de repuesto 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 147: ...ES APÉNDICE DIAGRAMAS MSA 74 ALTAIR 5 Manual de funcionamiento ...
Страница 207: ...FR INFORMATIONS POUR LES COMMANDES MSA 60 ALTAIR 5 Manuel d utilisation Pièces de rechange 8 1 7 11 2 5 6 3 4 12 10 9 ...
Страница 221: ...FR MSA 74 ALTAIR 5 Manuel d utilisation ...