Service setup
Page 52
-
Press the button
„Open CMD”
-
This is the command line for the operating-system. It
can be used to interact with the system.
Type “exit”
or click the sign “x” on the top
-right side of the
display to go back to the service setup
9.1.7
System Info. (system-information)
The settings in this page define the appropriate device type. Furthermore, e.g. Informations
of memory usage can be displayed.
This settings is only configurable for trained service technicians. Improper
operation or incorrectly stored configuration settings may lead to malfunction or
damage to the device
-
Press the button
„System Info.”
-
This is where the options can be adjusted
Used memory: total memory used (in bytes)
Available memory: total memory available (in bytes)
Start time: time and dates of the system start working
Serial Nr: serial number of the system (should be set once)
Device-Type & Version: device system type (should be set once)*
Desinf-Reminder: total time to remind user of system disinfection
Remainig days: total days until the next disinfection notification
UV-Lamp-Counter: how many times the UVPW has been activated
Option TOC: display TOC in the main screen (TI-System)
Option UF-Modul: activate the counter of UF-Modul (LS-System)
Option UV-Lamp: activate the counter of UV-lamp (AL-System)
Option normal start: if disabled, the next restart will show first start**
Option Debug: for system debugging with docklight
Option Simulation: to simulate system with constant values
*changing Device-Type will restart the system, so it is recommended to change the
Version before the Device-Type
**this option should be disabled, if the system should be restarted into FirstStart