Service setup
Page 50
-
Press the button “Q
-
” or “Q+“ to change the scale
-
value of the conductivity measurement (upper value)
or the offset (lower value)
-
Press the button “T+” or “T
-
“ to change the scale
-
value of the temperature measurement (upper value)
or the offset (lower value)
-
Entries change is saved automatically
The scale value is set in factory and should not be changed. The adjustment is
done by using the offset normally. For more detailed calibration instructions
please see chapter 9.2.1.
9.1.2
Limit setting of conductivity sensors
-
Press the button
„
Q-
Limits”
-
Press the button “
-
“ or “+” to shift the limit of
conductivity sensors in µS/cm
Note:
The unit of these conductivity limits is always
shown in µS/cm
9.1.3
Tank setup
-
Press the button
„Tank Setup”
-
Press the button
“V+” or “V
-
“ to change the maximum
voltage range of the tank level sensor
-
Press the button “VOL+” or “VOL
-
“ to change the
maximum scale-range of the tank level indicator.
-
Entries change is saved automatically