6.4
Replacement of the UV-lamp
Figure 16: UV-lamp
-
Follow the step in chapter 5.2.2.
-
Open the system.
-
Put the UV-reactor out of the clips.
-
Loosen the UV-lamp by unscrewing the union nut.
-
Remove the cap of the UV oxidator.
-
Unplug the UV lamp and remove it.
-
Insert the new UV-lamp carefully without touching the glass with your fingers.
-
Plug the cap on the new UV lamp.
-
Fix the UV-lamp by screwing the union nut.
-
Put the UV-reaktor into the clips.
-
Close the front door.
-
Reset the operation hours counter. Please follow the steps in chapter 5.2.2 or
manually in chapter 5.4.3 to do that.
Union nut
cap
UV-lamp
UV-reactor