5.5
Errors and Countermeasures
Display warning messages, their meanings and measures in order to eliminate them. If
these measure do not contribute to elimination contact the customer service.
Trouble
Reason
Measure
(1) Q3 will be incorrect
displayed! Change UF-
Module.
Ultrafiltration module
exceeded the maximum
operating hours.
Replace the ultrafiltration
module. Reset the expired
operating hours.
(2) OPERATING HOUR
LIMIT P1, Exchange
recommended
Reverse osmosis pump
exceeded the maximum
operating hours.
Inform the customer service.
(3) OPERATING HOUR
LIMIT C.MODUL,
Exchange necessary
Conditioning module
exceeded the maximum
operating hours.
Replace the conditioning module
Reset the expired operating
hours.
(4) OPERATING HOUR
LIMIT P2, Exchange
recommended
Circulation pump exceeded
the maximum operating
hours.
Inform the customer service.
(5) OPERATING HOUR
LIMIT UPW-PAK,
Exchange
recommended
Polisher module UPW-Pak
exceeded the maximum
operating hours.
Replace the polisher module
UPW-Pak. Reset the expired
operating hours.
(6) Please change UV
Lamp
UV-lamp exceeded the
maximum operating hours.
Replace the UV-lamp. Reset the
expired operating hours.
(7) OPERATING HOUR
LIMIT „UVT“, Exchange
recommended
UV-submersible-lamp
exceeded the maximum
operating hours.
Replace the UV-submersible-
lamp. Reset the expired operating
hours.
(8) Pump P1 error:
Check pressure
sensor!
No reverse osmosis-pressure
can be registered for more
than a half hour.
Check the main water supply.
Inform the customer service.
(10) Q2 is over limit!
Feed water quality is
insufficient. Reverse
osmosis function.
Conditioning module
exhausted.
Check pretreatment. Check
function of reverse osmosis.
Exchange conditioning module.
Check EDI-cell. Inform the
customer service.
(11) Q1 is over limit!
Check RO process.
Feed water quality is
insufficient. Pretreatment
module ProPak-R10
exhausted. Reverse osmosis
function.
Check pretreatment. Replace the
pretreatment module ProPak-
R10. Inform the customer service.
(12) Water level is too
low!
Product water tank is empty. Wait for tank filling. Check the
feed water supply. Check the