Press the button ”CHANGE ProPak”
or
p
ress the button “CHANGE C.Modul”
or
press the button “CHANGE UPW
-
Pak”
or
press the button “CHANGE UF
-
Modul”
*
(see also chapter 6.3 or chapter 6.6)
*
only for system with UF-modul
-
Tank level should be more than 50% to change
UPW-Pak or UF-Modul. If tank level is less than 50%,
these tools abort automatically.
-
If the tank level is more than 50%, it will release
system pressure for few seconds
-
Change the requested Pak;
o
ProPak is the one with black arrow
o
C.Modul is the one with red arrow
o
UPW-Pak is the one with blue arrow
o
UF-Modul is the transparent tube
-
Then click “Confirm”
-
Waiting few minutes to flush the new installed Pak:
o
ProPak: 5’
o
C.Modul: 60’ (AQ²E) or 15’ (AQ²P)
o
UPW-
Pak: 5’
o
UF-
Modul: 5’
-
To reset the operation hours please enter the code
which is provided by
membra
Pure
GmbH
OR
-
This step could be skipped (by clicking “cancel”), but
the operation hours is not resetted to zero