20
Continuous Sampling
Important! Do not connect Waste Fluid Hose to a pressurised
system, as this will cause the LPA² to malfunction and could
cause internal leakage. The Waste Fluid Hose must be dis-
charged into a tank/ vessel vented to atmosphere.
To conserve battery life, the LPA² should be permanently connected
to the power adaptor when it is operated in the continuous sampling
mode.
4.1 Continuous Sampling – Basic Operation
1
See page 11, instructions 1 to 5 inclusive, to select the appropriate
LPA² settings.
On the Settings Screen, press 5 repeatedly until
CONTINUOUS
is
selected.
2
Test Option – Press 6 :
Then press the relevant key to switch between option selections:
1 Sets the
MINUTES BETWEEN TESTS
Press 1 then input the time in minutes that is required between
the end of a test and the beginning of a new test. Input a value
between 1 and 30000 followed by
←
RETURN
.
2 selects between
LOG EVERY TEST: ON
and
LOG EVERY TEST: OFF
Selecting
LOG EVERY TEST: OFF
will store none of the test results
in the LPA²’s memory.
4 used to input the Clean Alarm Level (ISO). Enter a value of 0
(zero) if not used.