Continuous Sampling
23
Selecting
CONFIRM CLEANLINESS LEVEL: ON
instructs the Analyser
to repeat the sampling cycle until the Clean alarm level has been
achieved in two consecutive samples, before the Completed sta-
tus is displayed. Selecting
CONFIRM CLEANLINESS LEVEL: OFF
permits the Clean alarm to be achieved only one time before the
Completed status is displayed.
4 Clean Alarm Level (ISO)
Press 4 then input desired Clean Alarm Level in the Code for-
mat Number/ Number/ Number – any code number combination
can be input, from code 5 to 24, example 10/9/5
←
RETURN
For continuous testing until the ISO Code is achieved, selected
ISO Format, as described under the previous section on page 11
“Analyser Settings” instruction 4, 1.
Testing will automatically continue until each of the three num-
bers in the Code have been achieved (or better).
5 Clean Alarm Level (NAS1638 / AS4059E-1)
Press 5 then input desired Clean Alarm Level, as a single Class
number in the range 2 to 12 inclusive, example 6
←
RETURN
For continuous testing until the (NAS1638 / AS4059E-1) Class
is achieved, select NAS Format or AS4059E Table 1, as de-
scribed in “Analyser Settings” page 11 instruction 4, 1. Test-
ing will automatically continue until the Class number has been
achieved at each of the five micron size ranges covered by NAS
1638 & AS4059E Table 1. Note: AS4059E-1 denotes Table 1
of the AS4059E standard.
6 Clean Alarm Level (AS4059E-2)
Press 6 then input desired Clean Alarm Level in the format
1A/2B/3C/4D/5E/6F in the following range: