22
Continuous Sampling
4.1.1 Continuous Sampling – with Clean Alarm Levels –
Alarm Mode 1
This operating mode is similar to the Basic Operation, but in this
mode the LPA² will stop testing when the specified clean alarm level
is achieved.
A status of
COMPLETED
is shown on the LCD when the specified
clean alarm level is achieved.
(For other Alarm Modes refer to page 30).
9
Follow the section “Analyser Settings” (page 11), instructions 1 to
5 inclusive, to select the appropriate LPA² settings.
On the Settings Screen, press 5 repeatedly until CONTINUOUS is
selected
10
Test Options 6
Press relevant key to switch between option selections
1 Enter
MINUTES BETWEEN TESTS
Press 1 then input the time in minutes that is required between
the end of a test and the beginning of a new test. Input a value
between 1 and 30000 followed by
←
RETURN
.
2 Selects between
LOG EVERY TEST: ON
and
LOG EVERY TEST: OFF
Selecting Log Every Test: Off will only store the results of the
test when the Clean alarm Level is achieved – this saves memory
space.
3 Confirm Cleanliness Level
Press 3 to turn
CONFIRM CLEANLINESS LEVEL
On and Off.