Precautions
5
1 Precautions
The default language setting for the LPA² Laser Particle Analyser is
English
. To change the language setting refer to page 12.
1.1 Battery
It is recommended that the LPA² be charged for a minimum of 24
hours prior to first use, to fully charge the internal battery.
1.2 Internal Cleaning
Do NOT clean the LPA² or Bottle Sampler with Acetone or similar
solvents that are not compatible with the seals in the LPA². The
recommended cleaning fluid for internal flushing is Petroleum Ether
- see also Fault Finding, page 67.
The use of a 500μm coarse screen filter, screwed onto the HP con-
nector, is recommended for heavily contaminated systems. Refer to
page 64 for the part number.
1.3 LCD Visibility
If the LCD screen remains blank then refer to page 35 for recharging
instruction. To improve the LCD screen visibility, the Analyser can
be inclined by unlatching the two extension feet fitted to the bottom
of the case.