Battery Charging
35
10 Battery Charging
The LPA² is equipped with an internal rechargeable battery capa-
ble of sustaining 8 hours continuous operation following a 24 hour
charging period (approximately 100 tests).
To conserve battery power the LCD screen is illuminated at a re-
duced level when the LPA² is operated without an external power
supply connected.
When the Low battery level indicator
is displayed the LPA² re-
quires recharging as soon as possible.
BEFORE COMMENCING RECHARGING always press RED
button to switch off the Analyser.
To recharge, connect lead from power adaptor to the DC power in-
put socket on the LPA². Observe that the battery charging indicator
illuminates on the LPA².
Battery power may be conserved by:
•
Operating LPA² whilst connected to power adaptor
•
Switching LPA² OFF between samples
•
Turning Auto Print Mode OFF
Should the battery become completely discharged it is advisable to
allow a minimum of 15 minutes charge time prior to commencing a
test. The LPA² must remain connected to the power adaptor during
subsequent tests until the battery has had time to recharge.
Total discharge will shorten the battery lifetime so should be avoided
where possible.