Operation 2018-05-21
28
3.
After draining the regulator, turn the dial back to adjust the
pressure and push it back down.
Cleaning the Tray
Pull out the tray and clean it with a neutral detergent or
Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%) about
once a month.
♦
The tray cannot be autoclaved.
Autoclaving will cause it to rust.
♦
Do not use disinfectants that contain chlorine-based cleaners
or benzalkonium chloride or chlorinated aromatics.
These solutions will corrode the tray.
♦
If this tray is not cleaned, an oily film will collect on the
outside of the Lubrina.
Throw away the old filter and replace it with a new one.
Dispose of used filters according to regulations for medical
waste.
Do not fail to push the tray all the way in.
Otherwise, oil could leak out or oil mist might be released.
Tray
Air Regulator Dial