Operation 2017-12-27
20
Press the selector switch corresponding to the handpieces
which are connected.
Selector lamps light up.
(If you press the wrong one, just press it again to turn it off.)
Close the front door.
♦
The Lubrina will not start if the front door is not closed.
Press the Start Switch.
The handpiece lamps blink while maintenance is performed.
* Press the Start Switch again if the procedure was stopped
before finishing.
Do not select a coupling that does not have a handpiece
attached to it. This would result in oil spraying out when the
Start Switch is pressed.
Do not select a coupling that does not have a handpiece
attached to it. Otherwise, you may inhale sprayed oil mist,
or it may splash into your eyes.
Select the most appropriate lubrication mode for each
handpiece. Otherwise, too much oil will remain inside the
handpiece and it will spray out when using the handpiece.
Use this equipment with proper air pressure (0.3 MPa to 0.5
MPa). Otherwise, too much oil will remain inside the
handpiece and it will spray out when using the handpiece.