Operation 2018-05-21
21
♦
Never remove the spray can during maintenance.
♦
Opening the front door will immediately stop the maintenance
procedure.
* If a can runs out during maintenance, the empty lamp was
start flashing, and the procedure will stop. Replace the
can and press the Start Switch to repeat the procedure
from the beginning.
When the maintenance for all the handpieces is completed,
their lamps will go out, a beep will sound and the Finished
Lamp will light up.
Open the front door and remove the handpieces.
Hold the handpiece with one hand and push up the release
ring with the other.
(for Morita couplings)
♦
Use Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%) to
wipe oil off the handpiece.
For micromotor attachments, hold the attachment with one
hand and press the release button with the other.
Take out the handpieces and close the door.
Take care not to cut your fingers on the front door or other
parts.
Release Ring
Release
Button
Finished Lamp