Operation 2018-05-21
16
5.
Operation
Handpiece Autoclave Sterilization
To prevent the spread of grave, life-threatening diseases like HIV and hepatitis B, autoclave handpieces after
performing regular maintenance using the Lubrina.
<Take Off>
<Clean>
(1) Take the handpiece off its tube.
(Remove the bur as well.)
(2) Clean the head with a brush and running water.
(3) Wipe the handpiece with gauze dampened with Ethanol
for Disinfection (Ethanol 70 vol% to 80 vol%).
Lubrina Procedure
< Clean and Lubricate >
Perform chuck maintenance for Morita ball bearing
handpieces.
Perform the maintenance procedure.
<Clean>
Remove excess oil by wiping with Ethanol for Disinfection
(Ethanol 70 vol% to 80 vol%).
<Autoclave>
Autoclave in a sterilization pouch.
♦
Some oil may remain inside the handpiece even after autoclaving. Refer to the user instructions for each
handpiece for proper storage. (Storing the handpiece in an upright position is recommended.)