background image

 

The Interface Solution Experts 

11

FDY

PC-Programmable Frequency

Transmitter and Display

Programming the Range Parameters

The 

Range

 tab permits you to set or capture the 

acceptable input range and establish range limits.

• Use 

the

 Lower Range Value

 cell to set the input 

value that will correspond to a 4mA output. In the 
same way, use the 

Upper Range Value

 to set the 

input value that corresponds to a 20mA output. 

• 

Reverse the output (causing the output to increase 
as the input decreases) by setting the 

Lower 

Range Value

 higher than the 

Upper Range Value

.

• The 

Capture Lower Range

 and 

Capture Upper 

Range 

buttons permit you to trim the upper and 

lower range of the sensor (if necessary). Just click 
the appropriate button to begin trimming.

Figure 8.

  Range Parameters Tab

NOTE:

If the Lower Range Value is set to 0Hz in the Input 

Range the Fail On Input Time-out feature will not be 

available, the Lower Range Value must be set to a 

value greater than zero to enable this feature (box 

must be checked to apply feature)  .

Low Pass Filter Enable: 

Reduces noise by fi ltering out 

high frequency signals.

NOTE:

Moore Industries recommends having the Low-Pass 

fi lter turned on when the upper input range value is 

less than 150Hz (especially on the 10VAC - 250VAC 

units).

Averaging Filter:

 Shows the number of readings 

the FDY averages together to provide the displayed 
reading.  This can be used to compensate for bent 
blades in a monitored turbine by setting the value to 
equal the number of blades on the turbine.  (i.e. if the 
turbine has four blades, set the 

Averaging Filter

 to 

4

).

Minimum Debounce Period: 

 Available only when 

contact closure is selected in measurement mode.  
This value eliminates noise from a contact closure 
input, but lowers the maximum measurable frequency.

After selecting the input parameters, continue 
by clicking the 

Range

 tab and proceeding to the  

Programming the Range Parameters

 section

.

You are given the option of scaling your process 
variable reading.  To do this, click the 

Enabled

 check 

box in the 

Scaling 

section.  Scaling allows you to take 

your process variable and manipulate the reading to a 
more customized range.

• In 

the 

Upper Scaling Value

 text box, enter the 

value you wish displayed when your input is at its 
full range.  Repeat the step for the value you wish 
displayed when your input is at its zero value and 
enter that into the 

Lower Scaling Value

 box

.

At the 

Scaled Mode Display Label

 box, enter the 

specifi c EGU (Engineering Unit) that you want as your 
display.

• The 

Range Limits

 section allows you to designate 

when the input will be considered bad (over or 
under range).  For example, if you have set the 

Lower Range Value

 for 20, you can set the 

Under 

Range

 for 10 and check the 

Enable

 box, allowing 

the input to decrease below 10Hz before an error 
is registered. In this case, the display shows 

Under/Over range

, and the output is in 

Failure 

Mode 

(see 

Programming Output Parameters).

Содержание FDY

Страница 1: ...FDY FDYNovember 2013 165 730 00E1 PC Programmable Frequency Transmitter and Display PC Programmable Frequency Transmitter and Display...

Страница 2: ...ters 9 Programming the Input Parameters 9 Programming the Range Parameters 10 Programming the Output Parameters 11 Programming the Trimming Parameters 12 Sensor Trimming 12 Output Trimming 13 Loop Tes...

Страница 3: ...eatures a large integral display that shows real time process status About this Manual Wherever you see a Note Caution or WARNING pay particular attention A Note provides information to help you in av...

Страница 4: ...kohms typical 1kHz for input 6V for 0 02 30AC model 125kohms for 10 250AC model Load Capability 508 ohms 24V Output Current Limiting 3 8mA and 21 4mA for input under range and over range 3 6mA and 23...

Страница 5: ...Hz Table 6 Period Input Parameters Resolution 1mV 1mV 1mV Range 0 01 1 5V 0 002 0 15V 0 002 0 15V Hysteresis Measured Period 40 sec to 50sec 40 sec to 2sec 40 sec to 2sec Input Range 100mV 30V DC Reso...

Страница 6: ...3 25 in 62mm 2 45 in 64mm 2 50 in 66mm 2 58 in 76mm 3 00 in 43mm 1 70 in 18mm 0 70 in FDY COM PS 0 PS 0 6027 8 HZ Figure 2 FDY Dimensions 65mm 2 55 in GND GND 1 2 NPT 1 2 NPT 102mm 4 00 in 129mm 5 06...

Страница 7: ...rrent configuration of the attached FDY 4 Measurement Mode The drop down menu selects which type of measurement the FDY will monitor 5 Variables Device Status Communications These sections display the...

Страница 8: ...ion Program 1 2 3 4 5 6 7 Default Factory Configurations The following are the default factory settings for your unit Input Measurement Mode Frequency 0 100Hz Input Hysteresis 10mV Timeout Period 10s...

Страница 9: ...ions Cable Communications Software Programming the FDY All operating parameters of the FDY are set using a PC a serial communication cable and Moore Industries Intelligent PC Configuration software Wh...

Страница 10: ...ed to set this to the default setting of 2V maximum is 2 5V For input voltages below 3V it is recommended to set this to 2 3 of the peak to peak p p input signal voltage plus the low state signal volt...

Страница 11: ...ed reading This can be used to compensate for bent blades in a monitored turbine by setting the value to equal the number of blades on the turbine i e if the turbine has four blades set the Averaging...

Страница 12: ...ox This reduces the effect of sudden spikes and helps stop unnecessary alarms by smoothing the reading Failure Mode Failure Mode lets you set the output of the transmitter when it is malfunctioning It...

Страница 13: ...attempt to trim the FDY ensure that the correct sensor is attached to the transmitter then go to the Trimming tab as shown at the left and 1 Select the User radio button under Trim Mode then answer Y...

Страница 14: ...e output will no longer be fixed at 4mA 4 Continue by clicking the Fix Current High button This should cause the FDY to output 20mA If the output is a different value enter it into the Output Current...

Страница 15: ...int of the transmitter and the transmitter will send out a 4mA signal or the selected output signal if the zero output is set at a value other than 4mA when this point is reached Likewise the point yo...

Страница 16: ...onfiguration program you are ready to download the configuration file Downloading the Configuration File After the configuration has been set to your satisfaction it can be downloaded to the unit by s...

Страница 17: ...signal wiring should be 2 inches Note Some of Moore Industries instruments can be classified as receivers IPT2 IPX2 etc and some can be classified as transmitters TRX TRY etc while some are both a rec...

Страница 18: ...eas requiring intrinsically safe instrumentation Refer to the Special Conditions outlined below before proceeding with installation Special Conditions of Use The following instructions must be adhered...

Страница 19: ...The Interface Solution Experts 19 FDY PC Programmable Frequency Transmitter and Display...

Страница 20: ...isions of the Directive Fulfilled by the Equipment Group II Category 1G EEx ia IIC T4 Notified Body for EC Type Examination LCIE Notified Body Number 0081 33 av du G n ral Leclerc 92260 Fontenay aux R...

Страница 21: ...EC Type Examination Certificate ISSeP 08 ATEX 033 X Technical Standards Used EN 60079 0 2006 Explosive atmospheres Part 0 Equipment General requirements EN 60079 1 2007 Explosive atmospheres Part 1 Eq...

Страница 22: ...IES FOR BREACH OF ANYWARRANTY CONCERNINGTHE GOODS OR SERVICES SHALL BE FOR THE COMPANY AT ITS OPTION TO REPAIR OR REPLACE THE GOODS OR SERVICES OR REFUND THE PURCHASE PRICE THE COMPANY SHALL IN NO EVE...

Отзывы: