Chapter 5
5-4
SPECTRAmax 340PC Microplate Spectrophotometer Operator’s Manual
number and contact Molecular Devices Technical Support or your local represen-
tative for assistance.
If the instrument is functioning normally when using SOFTmax PRO,
no errors should be in the buffer (except error number 100).
NOTE:
Table 5.1: Error Codes, Messages, and Notes about the Errors
Error
Code
Error Message
Notes
100
command not found
Command string was not recognized by the instrument.
101
invalid argument
Command Argument does not match specified parameters or
syntax.
102
invalid data
Command argument is out of acceptable range.
103
too many arguments
Too many arguments after a specific command.
104
not enough arguments
Not enough arguments for a given command.
105
input line too long
Too many characters in the input line.
106
command error, system busy
Instrument could not perform the given command because it was
busy doing another task. Example: Request a wavelength while
the monochromator is in motion.
107
command error, measure-
ment in progress
Instrument could not perform the command because a measure-
ment was in progress.
108
command error, no data
available
A request to the instrument to transfer data when there’s no data
in the buffer.
109
data buffer full
Too many data sets in the buffer. Can be caused by setting up a
long kinetic run and disconnecting the computer or when
SOFTmax Pro is preempted by another application.
110
error buffer full
More than 65 errors in the buffer.
205
user generated bus error
Result of improper use of certain commands.
209
receive buffer overflow
Caused by an external device sending too much data over the
serial port and ignoring flow control.
210
serial port parity error
Parity bit error detected with incoming serial data.
211
serial port overrun error
Caused by host computer sending too much data and ignoring
the flow control signal.
300
temp sensor faulty
Unable to read a reasonable thermistor value. Thermistor is
faulty or disconnected, Main board problem, or ambient temper-
ature is out of range.
Содержание SPECTRAmax 340PC
Страница 4: ...iv SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 10: ...Chapter 1 1 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 30: ...Chapter 1 1 22 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 32: ...Chapter 2 2 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 36: ...Chapter 2 2 6 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 38: ...Chapter 3 3 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 44: ...Chapter 3 3 8 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 46: ...Chapter 4 4 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 52: ...Chapter 4 4 8 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 54: ...Chapter 5 5 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 58: ...Chapter 5 5 6 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 60: ...Appendix A A 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 62: ...Appendix B B 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 68: ...ii SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...