Maintenance
SPECTRAmax 340PC Microplate Spectrophotometer Operator’s Manual
4-7
7) It is possible that only one of the fuses may have blown. Molecular Devices
recommends that you replace both fuses, however, to ensure continued
proper operation. Pull both fuses out of the holder and discard them.
Figure 4.4
: The Fuse Box and Holder with Fuses Removed
8) Insert new SLOWBLOW-rated fuses into the fuse holder. Either end of the
fuse may be forward.
9) Insert the fuse holder into the fuse box, making sure that the fuses face
toward the right (toward the tongue on the cover) as you insert it. Slide the
fuse holder all the way into the box.
10) Insert the fuse box into the opening in the instrument, making sure that the
fuses are on the right side (toward the power receptacle). Press the fuse box
into place, making sure the cover snaps closed.
11) Reconnect the power cord to the instrument and to the wall outlet and
reconnect other cables previously disconnected.
Long-Term
Shutdown
If you will not be using the SPECTRAmax 340PC for an extended period of
time, clean the
external
surfaces of the instrument and cover it with the SPEC-
TRAmax dust cover.
Moving the
SPECTRAmax
340PC
If you need to relocate the SPECTRAmax 340PC, follow these steps.
The SPECTRAmax 340PC weighs approximately 29 pounds
(13.2 kilograms). To avoid injury, it is recommended that two people lift
the instrument together, using proper lifting techniques.
1) Remove any microplate from the drawer then close the drawer.
2) Turn off the power switch and unplug the power cord from the source and
from the receptacle on the back of the instrument.
3) Depending on the distance that you will be moving the instrument, you may
wish to repackage the SPECTRAmax 340PC in its original shipping carton.
Otherwise, carry the instrument or place it on a rolling cart to transport it.
4) Ensure that the new location meets the proper specifications as described in
Chapter 2, “Installation.”
WARNING:
Содержание SPECTRAmax 340PC
Страница 4: ...iv SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 10: ...Chapter 1 1 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 30: ...Chapter 1 1 22 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 32: ...Chapter 2 2 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 36: ...Chapter 2 2 6 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 38: ...Chapter 3 3 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 44: ...Chapter 3 3 8 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 46: ...Chapter 4 4 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 52: ...Chapter 4 4 8 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 54: ...Chapter 5 5 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 58: ...Chapter 5 5 6 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 60: ...Appendix A A 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 62: ...Appendix B B 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Страница 68: ...ii SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...