Mi Microplate Washer User Guide
88
5032022 C
Teach-In the Height for Aspiration
Use the "teach-in" method to define the height for aspiration.
Figure 7-12: The Aspiration Height "teach-in" prompt
Button
Action
On the left
With
displayed on the left, press this button to detect and move the aspiration
needle to the bottom of the well.
With
displayed on the left, press this button to move the manifold up.
On the right
Press this button to move the microplate to the currently saved aspiration position.
Press this button to increase the offset value by 0.1 mm to a maximum of 18.8 mm.
Press this button to decrease the offset value by 0.1 mm to a minimum of 0.0 mm.
Press this button to accept the displayed offset value and then define the dispense
position. See
Define the Dispense Position on page 89
.
Table 7-11: Actions of Control Panel Buttons