Confirm the Level of the Waste Bottle
If the
Waste Warning
option is enabled in the
Setup
menu, then the software displays a
prompt to make sure that the waste bottle has sufficient room for aspirated liquid during the
prime or rinse process.
CAUTION!
Running a prime or rinse process without sufficient room in the waste
bottle can damage the instrument. The minimum volume required for a Prime/Rinse
function is 7.5mL+1.5mL (total of 9mL), where 7.5mL is the volume of liquid in the tube
(from Rinse bottle to Dispense head of manifold) and 1.5mL is the volume dispensed
during the Prime/Rinse cycle. We suggest using a little more than 7.5mL + 1.5mL to
have enough backup pressure to push liquid through the tube and dispense the
required volume (1.5mL).
To enable the
Waste Warning
option in the
Setup
menu, see
.
Figure 4-20: The Waste Empty prompt
Button
Action
On the left
If the waste bottle is not empty, press this button to return to
Prime/Rinse
. See
Priming or Rinsing the Fluidics System on page 31
On the right
If the waste bottle is empty, press this button to define the volume of liquid to use for
the automatic prime or rinse process. See
Defining the Liquid Volume for the Prime
.
The button has no function for this option.
The button has no function for this option.
The button has no function for this option.
Table 4-13: Actions of Control Panel Buttons
Содержание MultiWash+
Страница 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Страница 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Страница 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Страница 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Страница 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...