32
Disconnect the electric power and the water supply to the system.
Remove the back access panel to the system.
Remove the heater bundle access plate.
Disconnect the heater bundle wires.
Remove the canister bolts.
Pull the heater bundle away from the canister. The S.S. diverter is
inserted between the heater elements. Note the position of the
diverter relative to the elements; it must be inserted in the
same position for the new heater bundle.
Clean the canister. Pay particular attention to the flanged area
where the gasket sits.
Insert the diverter between the heater elements of the new heater
bundle as follows (see figure to the right):
o
Locate the "N" stamped on the lead end of the heater bundle.
o
Set the heater bundle upright on the lead end with the "N"
facing right (3 o'clock position).
o
Hold the diverter upright over the heater bundle with the gap
on the top and facing right (3 o'clock position).
o
Slip the diverter straight down between the heater elements
so that each section has an equal number of heater elements.
Insert the new heater bundle into the canister with a new gasket so that the "N" stamped
on the lead end is in the 5 o'clock position.
(See figure below)
Replace the canister bolts.
Reconnect the heater bundle wires as diagramed on the back of the access plate.
Replace the access plate.
GAP
HEATER ELEMENTS
LEAD
END
Содержание Duratherm
Страница 56: ...51 Section 9 Model Codes See Following Pages...