Chapter Two: Installation
Setup
16
Mounting Screw Length
Note
The maximum screw penetration into the valve mounting flange is equal to the nominal screw
diameter. The use of screws exceeding this depth may result in improper flange preload and/or
damage to the valve.
Table 9: Mounting Screw Length
Nominal Size
Screw Quantity X Diameter
mm inch
ISO-F
JIS
320
250
200
160
100
12
10
8
6
4
NA
12 X M10
12 X M10
8 X M10
8 X M8
12 X M12
12 X M12
8 X M12
8 X M10
NA
Installation Procedure
1.
Install the control valve on the vacuum chamber. Chamber mounting side is indicated on the outline
drawing and is the side opposite the motor.
2.
Connect compressed air: Connections are for 1/4” tube
Use only clean, dry air or N2
Compressed air pressure (refer to Table 10 below)
Connect compressed air to controller connection “AIR/N2 INPUT”
3.
Connect the heating device (see
Appendix A: Product Specifications
4.
Option: Connect vacuum line to intermediate pumping port (1/8” NPT) rotary shaft seals.
Table 10: Required Air Pressure Based on Valve Size
Valve Size
Compressed Air Pressure
mm inch
psig
320
250
200
160
100
12
10
8
6
4
75-100
65-100
55-100
55-100
55-100
Содержание T3PIA
Страница 3: ...MKS Type T3PIA Pendulum Valve With Analog TTL Interface...
Страница 8: ...viii This page intentionally left blank...
Страница 27: ...Electrical Information Chapter Two Installation 19 Figure 4 Top Panel of the T3P Unit...
Страница 32: ...Chapter Two Installation Startup 24 This page intentionally left blank...
Страница 60: ...Appendix C Dimensions 52...
Страница 61: ...Appendix C Dimensions 53...
Страница 62: ...Appendix C Dimensions 54...
Страница 63: ...Appendix C Dimensions 55...
Страница 64: ...Appendix C Dimensions 56...
Страница 65: ...Appendix C Dimensions 57...
Страница 66: ...Appendix C Dimensions 58...
Страница 67: ...Appendix C Dimensions 59...
Страница 68: ...Appendix C Dimensions 60...
Страница 70: ...Index 62...