How To Check the Transducer Zero
Chapter Two: Installation
42
How To Zero an Absolute Transducer
To zero an absolute transducer, you must pump the unit, with the power on, down to a pressure
less than the transducer’s resolution (0.01% of Full Scale).
Note
The zero adjustment
must
be made at a pressure less than the transducer’s
resolution (0.01% of F.S.).
In addition, you should position the transducer in the
same orientation
as
it will be positioned when installed in your system.
Zeroing a transducer at a pressure above its stated minimum resolution creates a
zero offset
relative to true absolute pressure. All subsequent readings are then linear and accurate
relative to
the offset value
.
Note
If your system cannot achieve a sufficiently low pressure to set the
transducer zero, you may use a vacuum leak detector with sufficient
vacuum pumping (to achieve proper zeroing pressures). In this case,
mount the transducer on the leak detector
in the same plane of orientation
as it will be during actual use
.
To properly zero an absolute transducer, follow this procedure:
1.
Install the transducer in a system and connect a power supply/readout.
2.
Pump the system down to a pressure below the resolution of the transducer.
Refer to Table 18, page 43, for recommended pressure levels.
3.
Using a small screwdriver, adjust the
ZERO
pot until the readout displays zero (0000).
Refer to Figure 12, page 41, for the location of the
ZERO
pot.
Содержание Baratron 740C
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 55: ...Chapter Two Installation How To Check the Transducer Zero 45...
Страница 58: ......
Страница 60: ...General Chapter Four Operation 50 This page intentionally left blank...
Страница 72: ...How To Configure the Trip Points Chapter Five Type R7xxC Relay System 62 This page intentionally left blank...
Страница 82: ...General Description Appendix A Product Specifications 72 This page intentionally left blank...