Setup
Chapter Two: Installation
26
Setup
Mounting the Unit
The 7xxC Series unit can be mounted with the cylindrical end in either a vertical or horizontal
position. The mounting requirements allow any foreign matter entering the pressure port to fall
away from
the sensing diaphragm.
Mounting the Unit in a Vertical Position
Mount a unit with high full scale pressure (greater than or equal to 20 psi) in any position
vertically. If the unit has a low full scale pressure (less than 20 psi) and is used with gases that
contain particulates, mount it in a vertical position with the cylindrical end of the unit pointing
upward, as shown in Figure 3, page 26. Do not install such a unit with the cylindrical end of the
unit pointing downward because particulate impurities may accumulate on the sensing diaphragm
and alter the pressure reading.
Figure 3: Acceptable Vertical Mounting Positions
High Pressure
Flow-Through
Design shown
with 4-VCR
male fittings
High Pressure
Single-Ended
Design shown
with
1
/
2
inch
tube fitting
MKS
INSTRUMEN
ANDOVER, MA
MKS
INSTRUMEN
ANDOVER, MA
Содержание Baratron 740C
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 55: ...Chapter Two Installation How To Check the Transducer Zero 45...
Страница 58: ......
Страница 60: ...General Chapter Four Operation 50 This page intentionally left blank...
Страница 72: ...How To Configure the Trip Points Chapter Five Type R7xxC Relay System 62 This page intentionally left blank...
Страница 82: ...General Description Appendix A Product Specifications 72 This page intentionally left blank...