How To Connect to a Type PDR Series Readout
Chapter Two: Installation
40
Bendix Connector (option 1 pinout)
The correct cable to use for the Bendix connector is CB700-4, which has the mating Bendix
connector on the transducer end and flying leads on the PDR end. Table 10 lists the pin
assignments for the various colored wires on the flying leads end of the cable.
Connections Between a PDR Readout and a Transducer (Bendix Connector)
Transducer Signal
Flying Lead Color Code
(CB700-4)
PDR Signal
Pressure Output
Green
Pressure Input
+ Power Input
Red
+ 15 VDC
Pressure Return
White
Signal Ground (A GND)
Power Return
Black
Power Ground (D GND)
Table 17: Connections Between a PDR Readout and a Transducer (Bendix Connector)