Chapter Two: Installation
How To Connect to a Type PDR Series Readout
39
How To Connect to a Type PDR Series Readout
This section provides information on how to wire a Type PDR Series readout to a Type 7xxC
transducer with either a Type “D” or Bendix connector.
Caution
DO NOT use the -15 VDC output of the Type PDR readout to power
the 7xxC/800B Series transducer. When the -15 VDC signal of the
PDR is connected to the power return of the transducer, a short
between the -15 VDC of the transducer and the A GND of the PDR
occurs. The PDR readout will blank out. Either unit may be
damaged.
Note
This information replaces the color code information found in the PDR
manuals.
Type “D” Connector (either 9-pin or 15-pin High Density)
The correct cable to use for the 9 pin Type “D” connector is CB700-2; the cable for 15-pin High
Density connector is CB700-6. The correct cable has the mating Type “D” connector on the
transducer end and flying leads on the PDR end. Table 16 lists the pin assignments for the
various colored wires on the flying leads end of the cable.
Connections Between a PDR Readout and a Transducer (Type “D” Connector)
Transducer Signal
Flying Lead Color Code
(CB700-6 and CB700-2)
PDR Signal
Pressure Output
Red
Pressure Input
+ Power Input
Green
+ 15 VDC
Pressure Return
Black
Signal Ground (A GND)
Power Return
White
Power Ground (D GND)
Table 16: Connections Between PDR Readout and a Transducer (Type “D” Connector)
Содержание Baratron 740C
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 55: ...Chapter Two Installation How To Check the Transducer Zero 45...
Страница 58: ......
Страница 60: ...General Chapter Four Operation 50 This page intentionally left blank...
Страница 72: ...How To Configure the Trip Points Chapter Five Type R7xxC Relay System 62 This page intentionally left blank...
Страница 82: ...General Description Appendix A Product Specifications 72 This page intentionally left blank...