2.3 Setting-up
1.Install the Micrometer Head on the XY stage.
Loosen the hex-socket head screw on the bracket. Fully insert the stem of the Micrometer Head . Then,
tighten the hex-socket head screw with the stem in place.
The clamp screw is provided to secure the spindle of the Micrometer Head. If using a Micrometer Head
with a fitting hole in the stem, insert the stem so the hole is aligned with the clamp screw. If this puts the
scale on the Micrometer head in a poor position for viewing the zero graduation, adjust the scale position
by turning the Micrometer Head sleeve. If using a Micrometer Head with no hole to clamp the spindle,
lightly tighten the clamp screw.
2. Connect the surface illuminator cable to the connector on the power panel, which is at the back of the
Measuring Microscope. Insert the cable firmly into the connector and secure it with the nut.
3. Set the voltage selector plug to the supply voltage as follows.
XY stage
Clamp screw
Bracket
Hex-socket head screw
When the voltage selector on the back panel is as in the
upper left figure, the combination of two slide switches on
the upper and lower sides performs a voltage setup. Set up
the voltage according to the voltage setting method display.
When the voltage selector on the back panel is as in the
lower left figure, remove the fuse holder by turning it
counterclockwise using a screw driver.
Turn the voltage selector with a coin, etc., and align the
rated voltage display with the position of the arrow. Replace
the fuse holder as before.
Voltage setting
method display.
Voltage selector
Fuse holder
Voltage selector
2. Installation and setup
12
Содержание TM-500 Series
Страница 2: ......
Страница 8: ...CONTENTS vi ...
Страница 18: ...2 Installation and setup 16 ...
Страница 32: ...4 Maintenance 30 ...
Страница 36: ...5 Troubleshooting 34 ...
Страница 41: ... Reference Diagrams 㪏 Top view of XY stage TM 505 505R Detail drawing of T groove 8 Reference Diagrams 39 ...
Страница 42: ...8 Reference Diagrams 40 TM 510 510R Detail drawing of T groove ...