3.1.5 Replacing reticles
To replace reticles, use the supplied reticle setting screw, as described below.
1. Pull up and remove the eyepiece.
2. Screw the reticle setting screw into the reticle. Pull out the reticle setting screw along with the
reticle.
3. Screw the reticle setting screw into the reticle to be mounted. Insert the reticle into the eyepiece
mount.
The reticle is provided with a positioning pin. Fit the pin in the groove on the eyepiece mount and
insert the eyepiece as far as possible.
4. Remove the reticle setting screw and insert the eyepiece.
Reticle setting screw
Reticle setting screw
Reticle
Reticle
Positioning pin
Positioning pin
3.1.6 Adjusting the diopter
While looking into the eyepiece, turn the diopter adjustment ring until the reticle can be seen
sharply.