E-SOUM-PLUS Rev E
7
Recommended Separation Distances Between Portable And Mobile RF
Communications Equipment And The SonicOne CLINIC System
(Table
206)
The SonicOne CLINIC system is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the SonicOne CLINIC System can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
SonicOne
CLINIC system below, according to the maximum output power of the communications equipment.
Rated maximum output
power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2,7 GHz
d = 2.4√P
0.01
0.12
0.12
0.24
0.1
0.38
0.38
0.76
1
1.2
1.2
2.4
10
3.8
3.8
7.6
100
12
12
24
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Table 1.5 Recommended separation distances (EN table 206)
1.2.
Electrical Safety Statement
The SonicOne CLINIC system is designed and tested to comply with UL 60601-1 and EN 60601-1.
WARNING 1.3
The SonicOne CLINIC system generates high voltages within the console itself and the connected handpiece.
To avoid injury, the console should never be operated before ensuring that its cover is properly closed and
not tampered with. Do not attempt to remove or disassemble the cover. There are no user-serviceable parts
inside the console. All service should only be performed by an authorized Misonix representative. No
modification of this equipment is required.
WARNING 1.4 Proper system grounding can only be ensured when an approved, hospital-grade receptacle and matching
power cord are used. To avoid the risk of electric shock, this equipment must only be connected to a supply
with protective earth. Install plug and receptacles as per local regulations before operating the unit. Power
cord, plug and receptacle should be examined to verify that they are in good working condition before
connecting the console. Never pull on the power cord to remove it from the receptacle.
WARNING 1.5 Connecting the console to a power outlet with inadequate voltage or frequency may cause the unit to
malfunction or to create a shock or fire hazard. Confirm that the voltage selector switch on the console rear
panel is set to the local voltage setting and ensure that the correct fuses are being used. Refer to section
12.1 on instructions for fuse replacement.