E-SOUM-PLUS Rev E
49
12.2
Pump Head Replacement
The pump head may not be connected to the unit for shipping purposes.
Mount Pump Head
Position Pump Head
Shaft recess and bayonet fitting on
pump head rear
Pump drive shaft on console front
Turn pump head 45°clockwise.
Align drive shaft on console front and
shaft recess on pump head rear. Drive
shaft and recess must engage easily.
Rotate pump head slightly back and
forth to check engagement.
Lock pump head in place
Turn pump head clockwise until it locks in place. Arrow should be in the vertical
position pointing down.
Table 12.2.1 Assembly of pump head
Remove Pump Head
Disable Ultrasound
Switch to Standby Mode using enable/standby button. Confirm that Amplitude
setting is HOLLOW GREY.
Switch Console OFF
Switch console OFF and disconnect power cord.
Remove Tubing
Open pump cover. Remove tubing. Close pump cover.
Release Pump Head
Press and hold lock lever
on pump head bottom.
Turn pump head 45° counter clockwise.
Remove pump head
Pull pump head away from console until pump drive shaft clears. Release lock
lever.
Table 12.2.2 Disassembly of pump head