E-SOUM-PLUS Rev E
29
The console features air vents on the bottom. When installing the unit, ensure that these vents are not blocked in a way that
would prevent the circulation of air around the unit.
Figure 7.1.2 Underside view of console with air vents
NOTE 7.1
Adequate air circulation is needed to cool electronic components inside of the unit. Do not block the
cooling fan at the console rear or the air vents on the console bottom. Do not place the unit on a towel,
foam or other soft surface since the material may block the air vents. Blocking these vents may cause unit to
overheat and malfunction or create a shock hazard. A clear drape can be used to protect the console front
panel but do not cover the pump housing or other console portions.
7.2.
Console Set-up – Part I (Non-sterile)
Console Set-up | Part I
Switch Mains Power OFF
Set Mains Power switch on console rear to OFF.
Connect IV-pole
Connect IV-pole to receptacle in console rear.
Hang container with sterile physiological saline irrigant into IV-pole hook.
Irrigation tubing features IV-spike and is compatible with rigid bottles or flexible bags.
Check Mains Fuses
See section 12.1
Check Voltage Selector
Voltage selector switch (see section 6.1 for location) shall be placed in a position that is
dependent on the model and hospital voltage.
Model
Line Voltage
Selector Switch Position
E-SOP-GEN
120 VAC, 60 Hz
“115V”
220/230/240 VAC, 50/60 Hz
“230V”
Connect Electrical Power
Connect power cord to receptacle on console rear and to wall outlet.
Connect Footswitch
Connect footswitch cable to receptacle on console rear.
Footswitch connector and receptacle are keyed to ensure proper connection. Turn
cable connector until keys match. Insert connector fully into receptacle. Turn outer
connector ring clockwise to lock into position.
Footswitch may be covered with clear drape during Clinical use.
Switch Mains Power ON
Set Mains Power switch on console rear to ON.
Front panel will display Main Screen upon completion of system start.
Table 7.2 Console set-up - part I