![Misonix sonicone Clinic Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/misonix/sonicone-clinic/sonicone-clinic_instructions-for-use-manual_1806347048.webp)
E-SOUM-PLUS Rev E
48
Model BCM-GN-100V
Fuse Specifications
Line Voltage
Manufacturer
Manufacturer P/N
Rating
Description
100 VAC, 50/60 Hz
Cooper/Bussman
GDB-6.3
250V @ 6.3 A
Fast Acting, Low Break-
ing
200 VAC, 50/60 Hz
Cooper/Bussman
GDB-3.15
250V @ 3.15 A
Fast Acting, Low Break-
ing
Table 12.2 Console fuse specifications model
Fuse Replacement (The fuse holder is located on the console rear)
Disable Ultrasound
Switch to Standby Mode using enable/standby button. Confirm that Amplitude
setting is HOLLOW GREY.
Switch Console OFF
Switch console OFF and disconnect power cord.
Remove Fuse Holder
Pinch tab on fuse holder.
Pull fuse holder out.
Replace Fuses
Replace both fuses as specified above.
Mount Fuse Holder
Push fuse holder back into receptacle.
Switch Console ON
Connect power cord and switch console ON
Check Function
Confirm that console powers up and that Main Settings respond to activation of
buttons A-F.
Table 12.1.2 Fuse replacement