E-SOUM-PLUS Rev E
38
9.2.
Cleaning
Misonix Inc. has validated the cleaning procedures outlined below.
Misonix continually updates its sterilization and cleaning instructions as required. For the latest instructions and reuse
recommendations please contact your local Misonix representative.
WARNING 9.2
All reusable handpiece parts and accessories must be properly decontaminated, cleaned and sterilized
before each use as per instructions contained in this manual. Failure to do so may lead to infections, which
can ultimately cause patient death.
WARNING 9.3
Misonix Inc. has validated all cleaning and sterilization cycles given in this manual. It is highly recommended
that the procedures given in this manual for cleaning and sterilizing the SonicOne CLINIC System and related
accessories be followed. It is the responsibility of the user of this device or any accessories used with it to
validate procedures for cleaning and/or sterilization if they differ from the procedures as outlined in this
manual.
9.2.1
Manual Cleaning Procedure
The following items are considered reusable items and should be cleaned as recommended:
•
Console
•
Footswitch
•
Handpiece
•
Counter wrench
•
T-wrench
•
Probe cover
Probe, Probe Cover and Wrenches
Wash & Brush
•
Wash items with hot water mixed with an enzymatic detergent such as ASP Enzol® or Steris Pro-
lystica® according to standard hospital protocol. Follow manufacturer’s directions for preparing
solutions.
•
Probe cover and wrenches may be fully immersed.
•
Brush all passages at least four (4) times from FRONT to REAR, rotating the brushes during inser-
tion and inserting the brushes fully. This ensures clearing of debris from the internal passages.
•
Item’s exterior surface can be cleaned using a standard soft bristle cleaning brush.
Rinse
•
Rinse item under warm running water for a minimum of 1 minute to clear soap residue.
Dry
•
Dry item fully with absorbent towel or paper. Dispose of cloth or paper in accordance with Hospi-
tal or Clinic practices for contaminated wastes.
Table 9.2.1.1 Cleaning of probe cover and wrenches