E-SOUM-PLUS Rev E
41
9.3.
Sterilizing By Steam Autoclave
Sterilization Methods and terminology are based on ANSI/AAMI ST81 AND EN ISO 17664:2004 standards.
9.3.1
Reusable, Autoclavable Components
E-SOP-HP
Handpiece
E-PWRCH
T-Wrench
E-CWRCH
Handpiece/Counter Wrench
BCM-H2
Soft Tissue Probe Cover
9.3.1
Validated Steam Sterilization Cycles
SonicOne Clinic Handpiece With Sterilization Tray
SonicOne Clinic Handpiece Wrapped, No Tray
132°C (270°F)
134-137°C
(274-279°F)
134°C/273°F
Configuration
Items wrapped, NO TRAY
3
Items wrapped, NO TRAY
3
Items wrapped, NO TRAY
3
Cycle
Prevacuum
Prevacuum
Gravity
Preconditioning Pulses
4
4
None
Minimum Exposure
4 minutes*
3 minutes*
20 minutes
Minimum Dry Time
45 minutes
30 minutes
5 minutes
SonicOne Clinic Probes Wrapped, No Tray
Table 9.3.2 Steam sterilization cycles
132°C (270°F)
134-137°C
(274-279°F)
Configuration
Items placed in Misonix Sterilization
Tray E-SYSTRAY
1
or E-SYSTRAY-2
1
Items placed in Misonix Sterilization
Tray E-SYSTRAY
2
or E-SYSTRAY-2
2
Cycle
Prevacuum
Prevacuum
Preconditioning Pulses
4
4
Minimum Exposure Time
4 minutes*
3 minutes*
Minimum Dry Time
30 minutes
30 minutes