E-SOUM-PLUS Rev E
32
WARNING 7.1 Improper connection of the handpiece cable may present a shock hazard. Confirm that handpiece
connector is dry prior to plugging it in.
WARNING 7.2 Do not operate pump with pump cover in raised position. Rollers might pinch loose clothing or fingers.
Personal injuries may result. Cooling of the ultrasonic tip and irrigation at the treatment site will be
inhibited.
CAUTION 7.4
Do not place the soft silicone tube behind or in front of the rollers (latch removed in illustrations)
CAUTION 7.5
Do not pinch the soft silicone tube when the latch is locked.
CAUTION 7.6
Do not pinch barb fittings when closing the latch.
CAUTION 7.7
Prime the irrigation tubing prior to use. At all times ensure that the irrigant flows towards the handpiece
when footswitch is depressed. If no irrigant is flowing, cease use until flow is restored.
WARNING 7.3
Tip and irrigation temperatures may exceed the tissue necrosis point with SonicOne CLINIC accessories for
hard tissue removal if insufficient irrigation flow rates are used. Always set the irrigation flowrate for hard
tissue removal to a setting no less than the comparable vibration setting. For example, if the vibration
setting is 7, a minimum flow setting of 70% should be used.
WARNING 7.4 Heat is being generated at the tip/tissue interface. A continuous, lateral sweeping motion is recommended
for general bone/tissue removal in order to minimize contact duration with the ultrasonic tip and minimize
the temperature increase.
The SonicOne CLINIC system is now ready for the system check.